Since an HPLC column is a cylinder, the simplest estimate for the column volume is the equation V=L*pi*r2, where L = length of column (typically 50-250 mm, or 5-25 cm), and r=radius of the column, where typical internal diameters of HPLC columns are 2.1 mm, 3 mm, 4 mm, and 4.6 mm.

For example, suppose you have a column that is 25 cm long by 4.6 mm internal diameter (ID). Since the ID is in mm, you first convert to cm, then divide by 2 to get 0.23 cm radius. The column volume equation then is:

V = 25 * pi * (0.23)2 = 25 * pi * 0.0529 = 1.3225 * pi = 4.15 cm3

From there, you can convert cm3 to mL directly, so your column has a volume of 4.15 mL. However, you must also allow for the relative porosity of the packing material in your column, which is harder to measure. Typically, an unretained analyte will be injected through the column at a known flow rate, and the time it takes for the analyte to exit the column is used to determine a better approximation of column volume.

In the case of using an unretained analyte (which in reversed-phase HPLC, the analyte might be Uracil), using the same 25 cm by 4.6 mm column above and a 1 mL/min flow rate, suppose the analyte elutes from the column at 3.2 minutes. The column volume would then be 3.2 minutes * 1.0 mL/min = 3.2 mL, which does not agree with the calculated column volume. This is due to the fact that the particles in the column take up some of the volume of the column, so the total column volume is reduced by the amount of space they take up.

User Avatar

Wiki User

15y ago

What else can I help you with?

Related Questions

What is mean by delay volume in HPLC analysis?

Delay volume in HPLC analysis refers to the volume of liquid in the system that is not actively participating in the separation process. It includes the volume of tubing, fittings, and the void volume of the column. Minimizing the delay volume is important for maintaining good chromatographic resolution and reducing analysis time.


What is dead volum in hplc?

Dead volume in HPLC refers to the volume in the system that is not actively involved in the separation process. It includes the volume of tubing, fittings, and detector cell that the mobile phase passes through without interacting with the stationary phase or analytes. Dead volume can lead to band broadening and decreased resolution in chromatographic separations.


What is an HPLC column?

HPLC Column is one type of tube containing a stationary phase react with mobile phase to detect peak


How do you distinguised np-hplc and rp-hplc?

NP-HPLC (normal phase HPLC) separates compounds based on their polarity, where the stationary phase is polar and the mobile phase is nonpolar. RP-HPLC (reverse phase HPLC) separates compounds based on their hydrophobicity, where the stationary phase is nonpolar and the mobile phase is polar. RP-HPLC is more commonly used due to its versatility and ability to handle a wider range of compounds.


What is dead volume HPLC?

The dead volume in HPLC is 137.45. The dead volume in science is used in retention measurements and also in thermodynamic studies and the abbreviation HPLC stands for High Pressure Liquid Chromatography.


What can one read in the HPLC Column?

In an HPLC column one can see very small molecules such as ATP, histidine, glucose, uracil, and pyridine. It is a form high quality of liquid Chromatography.


What is void volume in hplc?

The void volume in HPLC is the volume of the column that is not occupied by the stationary phase. It represents the space where mobile phase flows through without interacting with the stationary phase or sample components. A large void volume can lead to poor resolution of peaks in chromatography.


What happen after used dipotassium buffer ph 9 on hplc column?

Nothing will happen


How do you calculate concentration from peak area in HPLC analysis?

To calculate concentration from peak area in HPLC analysis, you can use the formula: Concentration Peak Area / (Slope x Injection Volume). The peak area is obtained from the chromatogram, the slope is the calibration curve slope, and the injection volume is the volume of the sample injected into the HPLC system.


Retention time in Hplc?

Retention time in High Performance Liquid Chromatography (HPLC) refers to the time it takes for a compound to travel through the chromatography column and elute from the detector. It is a key parameter for identifying and characterizing compounds in a sample. Retention time is influenced by factors such as the column type, mobile phase composition, and compound properties.


What is the systm volume and what is the dead volume and what is the wide volume and what is the delay volme in HPLC?

system volume is amount of mobile phase travel from reservior to tip of the colomn . dead volume is the amount of mobile phase present in colomn. delay volume is amount of mobile phase used in at the time of injection and the time of detection of chromatogram.


Does HPLC can read histamine and TVB-N?

Yes, HPLC can be used to analyze histamine and TVB-N (Total Volatile Basic Nitrogen) in food samples. HPLC is a sensitive and selective technique that can separate and quantify various compounds, including histamine and TVB-N, based on their chemical properties. Pre-column derivatization may be required for some compounds to enhance their detection sensitivity in HPLC analysis.