Reagent Blank : Take reagent and add deionised water (in place of sample to be tested). Now measure the OD at specific wavelength --> this OD is your reagent blank. Substract this OD from your test result (with sample) to avoid any false +ve effect due to colour of reagents itself.

Sample Blank : Take sample and measure the OD without adding reagents --> this OD is your sample blank. Substract this OD from your test result to avoid any false +ve effect due to colour and turbidity of sample itself. As it is the fact that colour and turbidity of each sample would vary from one to another.

So now it is clear that Reagent blank is used to avoid bias due to colour of reagents and Sample blank is used to avoid bias due to sample itself.

User Avatar

Wiki User

11y ago

What else can I help you with?

Related Questions

If you want to read the absorbance at 595nm of a protein solution mixed with the colorizing solution called Bradford reagent what is a suitable reagent blank?

A suitable reagent blank for measuring the absorbance of a protein solution mixed with Bradford reagent at 595nm would be a blank containing all components of the reaction except the protein sample, such as water or buffer mixed with the Bradford reagent. This blank will account for any background absorbance contributed by the reagent itself, allowing for a more accurate measurement of the protein concentration.


What is a reagent blank?

A Reagent Blank contains the reagent(s) in the same concentration(s) and solvent(s) as would be contained in a sample prepared for analysis such that the Reagent Blank and a prepared sample only differ in that the Reagent Blank contains no sample and that none of the analyte(s) of interest has(have) been intentionally added to the blank. For samples containing radioactive isotopes that are prepared then analyzed by alpha particle, beta particle, or gamma ray counting techniques; analytical methods for the direct determination (no reagents are used) of the concentration of elements, ions, or compounds that absorb visible, ultraviolet, or infrared light, such as certain analyses by UV-Vis spectrophotometry and analysis by inductively-coupled plasma - atomic emission spectrophotometry (ICP-AES), ICP - mass spectrometry (ICP/MS) or atomic absorption (AA) spectrometry, preparatory blanks, often called "Prep Blanks," are used. These types of blanks are prepared exactly as if they were samples, for example by preparatory chromatography, solvent extraction, purge-and-trap methods, or acid or fusion digestion, but without any sample added. This type of blank is not technically a Reagent Blank, although it may sometimes be named as such. Reagent blanks and Prep blanks are used so that the contribution of any species not present, or not expected to be present, in the sample alone that adds to or subtracts from the detection signal may be evaluated or subtracted out (or added to) the detected concentration of the analyte.


Why do we add water to reagent blank?

The reagent blank should contain everything that the sample contains, except one variable. That variable could be the active ingredient, the enzyme, the substrate, or some other ingredient that is essential to the reaction. If water is added to all the other tubes, it must also be added to the reagent blank.


What is the difference between a blank monthly calendar and a regular calendar?

The difference between a blank monthly calendar and a regular calendar is that there are days of the weeks with numbers written on a regular calendar. A blank monthly calendar has nothing written on it, which you can fill in yourself.


What is the purpose of the blank?

Blank Sample in Spectrophotometry is used to measure the absorbance of light without sample. It is subtracted from the total absorbance for measurement of Absorbance from a sample's absorbance.


Fill in the blank The measure of an angle is the difference between the directions of its two?

sides


Fill in the blank. The measure of an angle is the difference between the directions of its two .?

sides


What is the purpose of blank samples?

Blank Sample in Spectrophotometry is used to measure the absorbance of light without sample. It is subtracted from the total absorbance for measurement of Absorbance from a sample's absorbance.


Difference between blank space and white space?

In computer programming, these two terms are interchangeable.


Why does blank titration shows the higher value compared to sample titration in iodine value?

In blank titration, no sample is present to react with the iodine solution, leading to an apparent excess of iodine. This can result in a higher value as all the iodine being counted towards the blank. In sample titration, the sample reacts with the iodine, leading to a lower amount of iodine available to react, resulting in a lower value compared to the blank titration.


Why do you use a reagent blank and not just distilled water to zero the spectrophotometer?

AS with any accurate machinery or scientific instrument there is a tendancy for them to get out of calibration over time, and after use. These products must meet specific requirements, codes, and standards. The purpose of "Zeroing" (or calibrating) is to make sure these items are brought back into "Zero" or "calibration" so that they can be as accurate as possible pertaining to the specifications that are required by the governing requirements, etc.. John Arndt/ CWI


Why is a blank solution used in morhs method?

A blank solution is used in the Mohs hardness test to serve as a point of reference for comparison. By observing how the blank solution interacts with the unknown mineral sample, the tester can accurately assess the hardness of the sample by determining if it scratches, is scratched by, or is comparable in hardness to the blank solution.